Digital Archives Initiative
Memorial University - Electronic Theses and Dissertations 1
menu off  add document to favorites : add page to favorites : reference url back to results : previous : next
 
 Search this object:
  
 0 hit(s) :: previous hit : next hit
  View:    
  previous page : next page
Document Description
TitleA computerized data acquisition and control system for Fabry-Perot inteferometry
AuthorTong, Shidong
DescriptionThesis (M.Sc.)--Memorial University of Newfoundland, 1990. Physics
Date1990
Paginationviii, 146 leaves : ill.
SubjectFabry-Perot interferometers; Interferometers
DegreeM.Sc.
Degree GrantorMemorial University of Newfoundland. Dept. of Physics
DisciplinePhysics and Physical Oceanography
LanguageEng
NotesBibliography: leaves 63-64.
AbstractFabry-Perot interferometry is one of the most powerful techniques in high resolution spectroscopy, especially in Raman and Brillouin scattering. Proper operation of a Fabry-Perot interferometer presumes the Fabry-Perot plates to be maintained in parallel alignment for a reasonable period of time while the experiment is in progress. Misalignment and laser frequency drift can cause very serious losses of finesse, and therefore of resolving power and contrast. -- Various methods of stabilizing the interferometer system have been reported previously. May et al. described a data acquisition and stabilization system for the Fabry-Perot interferometer which automatically compensates for all instrumental frequency drifts, maintains interferometer alignment, and permits extended data acquisition time in selected spectral regions of interest. The commercial version of this data acquisition system is known as DAS-1. It was originally produced by Burleigh Instruments, Inc. in 1976 and has been the accepted standard among commercially available instruments. However it has not been available for several years. -- Hardware and software packages which perform multichannel scaling (MCS) are now commercially available from several sources. The ACE-MCS package as supplied by EG&G-ORTEC and used in the present work consists of an IBM PC plug-in card and an MCS emulation program. This ACE-MCS package is simple and easy to use, but it lacks some essential features required in Fabry-Perot interferometry. The main problem is that it does not permit active feedback control of the interferometer alignment. In addition, it does not provide for a segmented ramp-voltage output, which is necessary in situations where the scattered light is very weak. -- In order to adapt this MCS package for Fabry-Perot interferometry, it was consequently necessary to modify and augment the hardware and software to accommodate the following functions: drift stabilization, finesse optimization, segmented ramp scanning. -- The drift stabilization and finesse optimization functions mainly required software development, while segmented scanning required hardware development in the form of a digital electronic circuit. -- A prototype system, based on an IBM-XT clone, has been developed at relatively modest cost and has been tested to record Brillouin scattering spectra of fused quartz. The experimental results show that the system is able to maintain Fabry-Perot alignment as well as the original Burleigh DAS-1 system.
TypeText
Resource TypeElectronic thesis or dissertation
FormatImage/jpeg; Application/pdf
SourcePaper copy kept in the Centre for Newfoundland Studies, Memorial University Libraries
Local Identifier76072876
RightsThe author retains copyright ownership and moral rights in this thesis. Neither the thesis nor substantial extracts from it may be printed or otherwise reproduced without the author's permission.
CollectionElectronic Theses and Dissertations
Scanning StatusCompleted
PDF File(13.79 MB) -- http://collections.mun.ca/PDFs/theses/ShidongTong.pdf
CONTENTdm file name51228.cpd